  2. Dual-Channel Off-Axis Ion Funnel With a Deflection Electrode

Dual-Channel Off-Axis Ion Funnel With a Deflection Electrode

  • Rapid Commun Mass Spectrom. 2025;39(20):e10103. doi: 10.1002/rcm.10103.
Mengfei Shan 1 La Chen 1 2 Jun Wang 1 Luhong Wen 1 2 3
Affiliations

Affiliations

  • 1 The Research Institute of Advanced Technologies, Ningbo University, Ningbo, China.
  • 2 China Innovation Instrument Co. Ltd., Ningbo, China.
  • 3 Hua Yue Enterprise Holdings Ltd., Guangzhou, China.
Abstract

Rationale: In electrospray ionization mass spectrometry (ESI-MS) systems, two critical challenges persist: (1) under-expanded supersonic jets at the atmospheric pressure interface (API) cause ion losses and reduced transmission efficiency; (2) residual Solvents and charged droplets entering vacuum stages lead to contamination and elevated chemical noise, degrading analysis accuracy.

Methods: A dual-channel off-axis ion funnel with a deflection electrode (DC-OFIDE) was developed to address these challenges. This device integrates three core components: an ion drift channel (IDC), an ion funnel channel (IFC), and a deflection electrode. The IDC and IFC are separated by conjoined gaps. Ions within the gas stream emanating from the API are extracted from the IDC via a deflection field, while a retarding axial field prolongs ions' residence time, ensuring efficient transfer to the IFC. This DC-OFIDE features an enlarged entrance aperture (Φ18 mm) to accommodate a multi-capillary interface, enhancing compatibility with high-conductance sample introduction systems.

Results: Compared with the original conventional ion funnel (CIF), the DC-OFIDE achieved a threefold enhancement in caffeine ion intensity and a broader m/z transmission window. It demonstrated robust neutral and droplet suppression, maintaining 80% ion intensity even under tripled serum volume infused. In drug screening of hair samples, baseline noises in drug ion peaks were reduced by 36%-82%, with a quadrupled signal-to-noise ratio improvement observed for 6-monoacetylmorphine.

Conclusions: This DC-OFIDE significantly enhances ion transmission efficiency and chemical noise suppression in ESI-MS, establishing its potential for high-fidelity analysis of complex samples.

Keywords

atmospheric pressure interface; contamination; dual‐channel off‐axis ion funnel with a deflection electrode (DC‐OFIDE); electrospray ionization mass spectrometry; ion mobility.